CS215 Temperature and Relative Humidity Probe
7. To change the SDI-12 address, enter the command aAb!; where a is the
current address from the above step and b is the new address. The CS215 will
change its address and the datalogger will exit the SDI-12 Transparent Mode.
8. To activate the SDI-12 Transparent Mode on Control Port p, enter pX and
press the <enter> key. Verify the new SDI-12 address by entering the ?!
command. The CS215 will respond with the new address.
9. To exit the SDI-12 Transparent Mode, enter ∗.
FIGURE 1. Screen Capture of SDI-12 Transparent Mode on Edlog Array-Based Datalogger (CR10X)
using Control Port 8 and Changing SDI-12 Address from 0 to 1
6.2 Table-Based “Edlog” Dataloggers
1. Connect a single CS215 to the datalogger Control Port p as described in
Section 4, and download a datalogger program that contains the SDI-12
Recorder (Instruction 105) instruction with valid entries for each parameter.
2. In the LoggerNet Toolbar, navigate to and activate the Test|Terminal
Emulator … menu. The “Terminal Emulator” window will open. In the Select
Device menu, located in the lower left hand side of the window, select the
station.
3. Click on the Open Terminal button. If communications between the
datalogger and PC are successful, the red bar located in the upper left hand
side of the window, will turn green.
4. Click inside the “Terminal Emulator” window and press the <enter> key until
the TD datalogger responds with the “>” prompt.
10