User manual

Table Of Contents
MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006
11.4 Returning unit
If the unit has to be returned, it should be carefully packed in the original packaging and returned to your authorised Gill
distributor, with a full description of the fault condition.
11.5 Status (error) codes
The Status code is sent as part of each wind measurement message. These can be observed on the RS232 programming
port inside of the junction box.
Code Status Condition
00 OK Sufficient samples in average period
01 Axis 1 failed Insufficient samples in average period on U axis
02 Axis 2 failed Insufficient samples in average period on V axis
04 Axis 1 and 2 failed Insufficient samples in average period on both axes
08 NVM error NVM checksum failed
09
10
ROM error
System gain at
maximum
ROM checksum failed
Wind velocity reported could be in error
A - NMEA data valid
V - NMEA data invalid
26