Specifications
Section 7. Measurement Instructions 
7-38 
Parameter 
& Data Type 
Enter 
CS7500Cmd 
The CS7500Cmd parameter requests the data to be retrieved from the 
sensor. The command is sent first to the device specified by the 
SDMAddress parameter. If the Reps parameter is greater than 1, subsequent 
CS7500s will be issued the command with each rep. The results for the 
command will be returned in the array specified by the Dest parameter. A 
numeric code is entered to request the data: 
 Code 
Description 
  0 
Get CO2 & H2O molar density (mmol/m3) 
  1 
Get CO2 & H2O absorptance 
  2 
Get internal pressure estimate (kPa), auxiliary measurement A, 
auxiliary measurement B, and cooler voltage (V)
  3 
Get cell diagnostic value, output bandwidth (Hz), and 
programmed delay [230 + (delay * 6.579)] (msec)
  4 
Get all data (CO2 molar density (mmol/m3), H2O molar density 
(mmol/m3), CO2 absorptance, H2O absorptance, internal 
pressure estimate (kPa), auxiliary measurement A, auxiliary 
measurement B, cooler voltage (V), cell diagnostic value, output 
bandwidth (Hz), and programmed delay [230 + (delay * 6.579)] 
(msec))
  5 
Get CO2 & H2O molar density (mmol/m3) and internal 
pressure estimate (kPa)
CSAT3 (Dest, Reps, SDMAddress, CSAT3Cmd, CSAT3Opt) 
Communicates with the CSAT3 three dimensional sonic anemometer. See 
CSAT3 manual for more information. 
Parameter 
& Data Type 
Enter 
Dest  The Dest parameter is a variable in which to store the results of the 
measurement. This variable must be dimensioned to a length of five to hold 
the CSAT3 Ux, Uy, Uz, speed of sound, and diagnostic data. 
Reps  The Reps parameter is the number of times the measurement should be 
made. Measurements are made on consecutive channels. If the Reps 
parameter is greater than 1, the Dest parameter must be a variable array. 
SDMAddress  The SDMAddress parameter defines the address of the CSAT3 with which 
to communicate. Valid SDM addresses are 0 through 14. Address 15 is 
reserved for the SDMTrigger instruction. If the Reps parameter is greater 
than 1, the datalogger will increment the SDM address for each subsequent 
CSAT3 that it communicates with. 
The SDM address is entered as a base 10 number, unlike older, jumper-
settable SDM instruments that used base 4. 