Specifications
Section 7. Measurement Instructions 
7-19 
7.8 Digital I/O 
PeriodAvg (Dest, Reps, Range, SEChan, Threshold, PAOption, Cycles, 
Timeout, Mult, Offset) 
This instruction measures the period of a signal on any single-ended input 
channel. The specified number of Cycles are timed with a resolution of 70 ns, 
making the resolution of the period measurement 70 ns divided by the number 
of Cycles chosen. 
Parameter 
& Data Type 
Enter 
Dest 
Variable or 
Array
The Variable in which to store the results of the instruction. When Reps are used the 
results are stored in an array with the variable name. An array must be dimensioned to 
have elements for all the Reps.
Reps 
Constant 
The number of repetitions for the instruction on adjacent channels. 
Range 
Constant 
The voltage range for the measurement, which determines the gain applied to the signal 
prior to a zero-crossing detector. Maximum frequency decreases with increasing gain.
Range Gain  Signal (pk-pk)
1
 Minimum Maximum 
Code  Min Max Pulse 
Width 
Frequency
2
mV5000  0.8  600 mV  10.0 V 
2.5 µs 
200 kHz 
mV1000  4.0  100 mV  2.0 V 
5.0 µs 
100 kHz 
mV200  20  4 mV  2.0 V 
25 µs 
20 kHz 
1
 Signals must cross threshold to trigger the voltage comparator. 
2
 Maximum frequency equals 1/(Twice Minimum Pulse Width) for 50% duty cycle 
signals. 
SEChan 
Constant 
The single-ended channel number on which to make the first measurement. If the channel 
is entered as a negative number, all reps occur on that channel.
Threshold 
Constant 
The voltage in millivolts that the input must cross for a count to occur. For a signal 
centered around CR5000 ground (Figure 7.8-1) the threshold should be 0. If the input 
signal is a 0 to 5 V CMOS signal then a threshold of 2500 mV would result in the voltage 
comparator switching at 2.5 V.
PAoption 
Specifies whether to output the period in µs or the frequency in Hz. 
 Numeric Code  Voltage Range 
0
  Period of the signal is returned 
1
  Frequency of the signal is returned 
Cycles 
Constant 
The number of cycles to be measured for the average calculation.  
Timeout 
Constant 
The maximum time duration (in msec) that the logger will wait for the number of Cycles 
to be measured for the average calculation. 
Mult, Offset 
Constant, 
Variable, 
Array, or 
Expression 
A multiplier and offset by which to scale the raw results of the measurement. 
Low-level signals are amplified prior to a voltage comparator for the period 
averaging measurement. The internal voltage comparator is referenced to the 
user entered threshold. The threshold parameter allows a user to reference the 
internal voltage comparator to voltages other than 0 V. For example, a 
threshold of 2500 mV allows a 0 to 5 V digital signal to be sensed by the 
internal comparator without the need of any additional input conditioning 
circuitry. The threshold allows direct connection of standard digital signals, but 