Specifications
Section 7. Measurement Instructions 
7-17 
Unless the AllRanges option is selected, the calibrate instruction only measures 
the range and integration combinations that occur in the measurements in the 
program. For the 250 µs and zero integration calibrations multiple 
measurements are averaged for the calibration values. The 250 µs integration 
calibration averages five measurements and the zero integration calibration 
averages ten measurements. 
The Calibration instruction can occur in a fast scan or in a slow sequence scan. 
In a fast scan the entire calibration is completed at once. In a slow sequence 
scan the calibration measurements are separated into sections that can be 
spliced on to the end of fast sequence scans. 
If it is necessary to update the calibration more rapidly than is done by the 
background calibration, try running the Calibrate instruction in the fast scan 
with the measurements. If there isn’t time for it to run there it can be placed in 
a slow sequence scan, but remember, unless the slow scan is faster than about 
40 seconds the calibration isn’t being updated any faster than with the 
background calibration. 
Running Calibrate in a slow sequence scan is not an option when there is not 
time for the automatic background calibration. The instruction requires more 
time because of the multiple measurements for the 250 µs and zero 
integrations. 
When the results of the calibration are placed in an array, the array must have 
at least 60 elements, more if the program contains instructions which use 
excitations. The calibration values will be in the following order, followed by 
the calibrations of the excitations if any. If a range is not calibrated, 0 will be 
returned for the gain and offset. 
Table 7.7-1. Calibrate Return Value Decode 
Array Element  Description 
1  zero integrate 5000 mV single ended offset 
2  zero integrate 5000 mV differential offset 
3  zero integrate 5000 mV gain 
4  zero integrate 1000 mV single ended offset 
5  zero integrate 1000 mV differential offset 
6  zero integrate 1000 mV gain 
7  zero integrate 200 mV single ended offset 
8  zero integrate 200 mV differential offset 
9  zero integrate 200 mV gain 
10  zero integrate 50 mV single ended offset 
11  zero integrate 50 mV differential offset 
12  zero integrate 50 mV gain 
13  zero integrate 20 mV single ended offset 
14  zero integrate 20 mV differential offset 
15  zero integrate 20 mV gain 
16 
250 µSec integrate 5000 mV single ended offset 
17 
250 µSec integrate 5000 mV differential offset 
18 
250 µSec integrate 5000 mV gain 
19 
250 µSec integrate 1000 mV single ended offset 










