Specifications
Section 7. Measurement Instructions 
7-9 
BrFull6W (Dest, Reps, Range1, Range2, DiffChan, ExChan, MeasPEx, 
ExmV, RevEx, RevDiff, SettlingTime, Integ, Mult, Offset) 
X = result w/mult = 1, offset = 0 
X
V
V
R
RR
R
RR
==
+
−
+
⎛
⎝
⎜
⎞
⎠
⎟
1000 1000
2
1
3
34
2
12
This Instruction applies an excitation voltage and makes two differential 
voltage measurements. The measurements are made on sequential channels. 
The result is the voltage measured on the second channel (V
2
) divided by the 
voltage measured on the first (V
1
). The result is 1000 times V
2 
/ V
1 
or 
millivolts output per volt of excitation.  The connections are made so that V
1
 is 
the measurement of the voltage drop across the full bridge, and V
2
 is the 
measurement of the bridge output. 
Full Bridge Parameters 
Parameter 
& Data Type 
Enter 
Dest 
Variable or 
Array
The Variable in which to store the results of the instruction. When Reps are used the 
results are stored in an array with the variable name. An array must be dimensioned to 
have elements for all the Reps. 
Reps 
Constant 
The number of repetitions for the measurement or instruction. 
Range 
Constant
Alpha Code 
Numeric 
Code 
Voltage 
Range 
  mV5000 
0  ± 5000 mV         
  mV1000 
1  ± 1000 mV         
  mV200 
2  ± 200 mV         
  mV50 
3  ± 50 mV         
  mV20 
4  ± 20 mV   
  AutoRange 
5  mV20 - mV5000  Selects range (Sect. 3.1) 
  mV200C 
mV50C 
mV20C 
20 
30 
40 
± 200 mV 
± 50 mV 
± 20 mV 
The mV200C, mV50C, and mV20C 
ranges pull the channel into common 
mode range and check for open input 
DiffChan 
Constant 
The differential channel number on which to make the first measurement. When Reps are 
used, subsequent measurements will be automatically made on the following differential 
channels. If the channel is entered as a negative number, all reps occur on that channel. 