Product manual

AM16/32B Relay Analog Multiplexer
analog ground. Up to sixteen full bridges may be multiplexed through the
AM16/32B.
A problem with making full bridge measurements with this configuration is
that the resistance of the lead wire and multiplexer relays can cause a voltage
drop, reducing the excitation at the bridge. The following section describes a
configuration that compensates for this by measuring the excitation at the
bridge.
6.5 Full Bridges with Excitation Compensation
Sensor to Multiplexer Wiring — With panel switch set to “4x16” mode, you
are 2 lines short for a six wire measurement. One solution is to multiplex the
four signal wires through the AM16/32B, but bypass the AM16/32B with
excitation and ground wires. This means that the sensors will be excited in
parallel which causes a higher current drain, possibly enough to exceed the
current available from the datalogger’s excitation channel. Alternatively, the
excitation an ground leads can be multiplexed through an additional
AM16/32B a owing the sensors to be excited one at a time (FIGURE 6-7). In
ultiplexer are
e multiplexer to
“8x16”.
D, EVEN
in the
d AM16/32B.
reset and clocked by the same control ports and/or
excitation channels to simplify programming.
d
ll
this case the 12V, GND, CLK, and RES lines of the second m
wired in parallel with those of the first, effectively widening th
Multiplexer to Datalogger Wiring Four leads from the COM OD
terminals connect to two sequential differential analog channels
datalogger. Excitation and ground are multiplexed by the secon
Both multiplexers can be
"4 X 16" Mode
COM H (ODD)
COM L
COM H (EVEN)
COM L
ODD H
ODD L
EVEN H
EVEN L
SENSOR SHIELDS
COM
"4 X 16" Mode
COM H (ODD)
COM L
ODD H
ODD L
/
0
CR23X/
CR800/
CR850/
CR1000
21X/
CR7
CR10(X)
CR3000/
CR5000
AG
EX E E VX
H H H H
L L L L
H H H H
L L L L
G
ensation FIGURE 6-7. Full bridge measurement with excitation comp
29