Instruction manual
9522B Iridium Satellite Modem
_____________________________________________________________________________
19
Figure 19: Generic - Modem Configuration
3.1.7. Step 7: PakBusPort Configuration
As in Figure 20, be sure the Communications Enabled box is
checked, change the Maximum Baud Rate to 19200 (9600 for a
CR10X), and add a 200 ms delay under Delay Hangup. You may
change other settings in this configuration to suit your particular
application. Once complete move to Step 8 in order to configure
the datalogger.
Figure 20: PakBus Port Configuration