SS100 Sensor Simulator 12/10 C o p y r i g h t © 2 0 0 7 - 2 0 1 0 C a m p b e l l S c i e n t i f i c , I n c .
Warranty and Assistance The SS100 SENSOR SIMULATOR is warranted by Campbell Scientific, Inc. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. Campbell Scientific, Inc.'s obligation under this warranty is limited to repairing or replacing (at Campbell Scientific, Inc.'s option) defective products.
SS100 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. General .........................................................................1 2. Specifications ..............................................................1 3. Wind Speed and Direction Simulation .......................1 4. Tipping Bucket Rain Gage Simulation.......................3 5.
SS100 Sensor Simulator 1. General The SS100 is a device used when the Campbell Scientific CR1000/LoggerNet Training Course is taught out-of-house. The SS100 is powered by a nominal 12VDC power source, typically the datalogger itself, and provides several outputs typical of those from sensors commonly measured by Campbell Scientific dataloggers. The Training Course instructors may use the SS100 in a variety of simulations of their own choosing or use the analogies in this manual. 2.
SS100 Sensor Simulator CRBasic Measurement Instruction: VoltSe (WindDir,1,mv2500C,1,1,0,_60Hz,360/2200,0) Wind Speed (Use SS100 High Frequency Pulse Output) Range: 0-50 m/s Accuracy: +/- 5% > 5 m/s or 0.2 m/s < 5 m/s Resolution: 0.1 m/s Wind Speed Output: Linear, Calm = 3Hz, 50 m/s = 140 Hz Wiring SS100 P G CR1000 P1 Multiplier and offset calculation: Using the equation of a line with two known points (0,3) and (50,140): y=mx + b; m = (50-0) / (140-3) = 0.365 m/s / Hz b=y-mx; 50 - .365(140) = -1.
SS100 Sensor Simulator 4. Tipping Bucket Rain Gage Simulation Sensor: Tipping Bucket Rain Gage Signal Output: Momentary switch closure activated by tipping bucket mechanism. Rainfall per tip: 0.01 in.
SS100 Sensor Simulator Wiring SS100 Vin V1out V2out G CR1000 EX1 2H 2L Multiplier and offset calculation: Results with mult = 1, offset = 0, minimum depth = -177, maximum = 500. Assuming the range of the sensor is 0-100 feet, calculate the multiplier and offset using equation of line; m = 0.15, offset = 26.14. CRBasic measurement instruction: BrFull (depth_ft,1,mv2500C,2,Vx1,1,2500,True ,True ,0,_60Hz,0.15,26.
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbellsci.com Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.csafrica.co.za • cleroux@csafrica.co.za Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 444 Thuringowa Central QLD 4812 AUSTRALIA www.campbellsci.com.au • info@campbellsci.com.au Campbell Scientific do Brazil Ltda.