Step 3: Position 1: Center position
Attaching the velocity calibration unit
Fix the PWC100 velocity calibrator unit to the center position on the PWS
using two of the countersunk fixing screws provided
Check that the velocity calibrator is located correctly and securely.
LED C should now be flashing slowly and all other LEDs should be off.
Once you are ready, press START.
The test is now automated and the disk should start rotating. LED C will
stay illuminated until this stage of the test is complete.
Wait for the calibration disk to stop rotating. This test will take
approximately two and a half minutes.
Upon successful completion of the test, LED L will be flashing slowly and
LED C will be turned off.