Manual
Recorded data can be verified from the keyboard/display in *7 mode.
Data retrieval should be done with a PC running PCPST.
07:XXXX where XXXX is the next location in the ring memory
where data will be stored.
Commands Specific to *7 Mode:
# Display Final Storage location number:enter location to
advance to, or C to display data
#A Advance to same element in next array with the same
ID
#B Back up to same element in previous array with the
same ID
Example:
Only one data point at a time can be viewed. The data will look some-
thing like this for the first 0.3 seconds of a Delta interval, 0.0 datum,
elapsed time test with 3 wells:
01:99 Indicates that the following is a header. This header
contains the startup information when a test is stopped.
02:1106 Start time (military time)
03:74.5 Starting depth of water
04:0003 Number of wells
05:0097 Location ID number (chosen by the user in *4 table)
06:0217 Day of year
07:12:29 Battery voltage at start of test
Then proceeding into the data
01:0001 Well number
02:.0075 Time in decimal minutes
03:.423 Water level
04:0002 Well number
05:00.875 Time
06:0.366 Water level
07:0003 Well number
ETC.
3.5 *7 Mode Dis-
playing Final
Readings
SECTION 3. TEST PROCEDURE
3-8