Instruction Manual

IRTS-P Precision Infrared Temperature Sensor
5
2: Thermocouple Temp (DIFF) (P14)
1: 1 Reps
2: 21 2.5 mV 60 Hz Rejection Range
3: 2 DIFF Channel
4: 3 Type K (Chromel-Alumel)
5: 1 Ref Temp (Deg. C) Loc [ RefTemp ]
6: 2 Loc [ AppTargT ]
7: 1.0 Mult
8: 0.0 Offset
;Measure Sensor Body Temp
3: Thermocouple Temp (DIFF) (P14)
1: 1 Reps
2: 21 2.5 mV 60 Hz Rejection Range
3: 3 DIFF Channel
4: 3 Type K (Chromel-Alumel)
5: 1 Ref Temp (Deg. C) Loc [ RefTemp ]
6: 3 Loc [ SenBodyT ]
7: 1.0 Mult
8: 0.0 Offset
;Calculate P, H, & K Coefficients
4: Polynomial (P55)
1: 1 Reps
2: 3 X Loc [ SenBodyT ]
3: 4 F(X) Loc [ Psb ]
4: 49.9092 C0
5: 0.59237 C1
6: 0.00558 C2
7: 0.0 C3
8: 0.0 C4
9: 0.0 C5
5: Polynomial (P55)
1: 1 Reps
2: 3 X Loc [ SenBodyT ]
3: 5 F(X) Loc [ Hsb ]
4: 4.2828 C0
5: 0.4248 C1
6: -0.00077 C2
7: 0.0 C3
8: 0.0 C4
9: 0.0 C5
6: Polynomial (P55)
1: 1 Reps
2: 3 X Loc [ SenBodyT ]
3: 6 F(X) Loc [ Ksb ]
4: 52.0705 C0
5: -5.3816 C1
6: 0.387 C2
7: 0.0 C3
8: 0.0 C4
9: 0.0 C5