Manual
HMP50 Temperature and Relative
Humidity Probe
1. General
The HMP50 Temperature and Relative Humidity probe contains a Platinum
Resistance Temperature detector (PRT) and a Vaisala INTERCAP® capacitive
relative humidity sensor.
The -L option on the model HMP50 Temperature and Relative Humidity probe
(HMP50-L) indicates that the cable length is user specified. Lead length is
specified when the sensor is ordered. Table 1 gives the recommended lead
length. This manual refers to the sensor as the HMP50.
TABLE 1. Recommended Lead Lengths
2 m Height Atop a tripod or tower via a 2 ft crossarm such as the CM202
Mast/Leg CM202 CM6 CM10 CM110 CM115 CM120 UT10 UT20 UT30
9' 11' 11' 14' 14' 19' 24' 14' 24' 37'
Note: Add two feet to the cable length if you are mounting the enclosure on the leg base of a light-weight tripod.
2. Specifications
Operating Temperature: -40°C to +60°C
Storage Temperature: -40°C to +80°C
Probe Length: 7.1 cm (2.8 in.)
Probe Body Diameter: 1.2 cm (0.47 in.)
Filter: 0.2 μm Teflon membrane
Filter Diameter: 1.2 cm (0.47 in.)
Housing Material: chrome-coated aluminum and
chrome-coated ABS plastic
Power Consumption: <2 mA
Supply Voltage: 7 to 28 VDC
Settling Time after power is switched on: 1 second
1