Owner manual
CS725 Snow Water Equivalency Sensor
The menus can be switched between English and French. To change the
language setting, key the command .L<Enter>. This will allow the selection of
either English or French. The English version of the Main menu is as follows:
*** LIST OF COMMANDS ****
[.l] Language (French or English)
[.h] Read the current live histogram
[.hp] Display the K and Tl peak positions from the live histogram
[.hl] Display the current histogram in log scale
[.hr] Display results after completionof a manual measurement (histogram,
position and #counts K and TI).
[.e] Same as <h> command , but erases histogram at end of operation
[.a] Special command: define start/stop time limits for one measurement
[.t] Display temperature data in °C
[.d] Display date and time
[.dd] Modify date and time
[.hd] Read back histograms from EPROM according to selected dates
[.f] Read back measurement results from EPROM according to dates
[.fl] Read back the current day measurement results
[.flla] Read back last calculated line result
[.fs] Read back calculated line result in short form ? (Time and EENS
only)
[.p] Display Parameter menu
[.s] Display status and current measurement information
[.0xps] Parameters save to file function
[.0xpr] Parameters recovery from file function
[%Rhard]/[%Rboot] Restart software processor/reboot firmware
[UU] CS725/terminal synchronizing sequence for rates other than 9600 bps
6.4 Detailed Main Menu Command Information
1. [.l]
The [.l] command is used to set the language setting to either French or
English.
The following menu is activated once the [.l] command is received:
[0] Francais
[1] English
Type the number 0 for French or the number 1 for English followed by
<Enter>.
5