User Manual

CS475, CS476, and CS477 Radar Water Level Sensor
TABLE 4-2. Radiation Beam Spread for CS475
(10° Beam Angle)
Distance in Meters Diameter of Footprint in Meters
1 0.18
5 0.87
10 1.76
15 2.64
20 3.53
TABLE 4-3. Radiation Beam Spread for CS476/CS477
(8° Beam Angle)
Distance in Meters Diameter of Footprint in Meters
1 0.14
5 0.70
10 1.41
15 2.11
20 2.81
30 4.216
70 (CS477 only) 9.84
3. Securely mount the sensor.
4. Use a user-supplied bubble level or the 25619 bubble level to make certain
the antenna horn is aligned within of vertical. The cap needs to be
removed when using the 25619. If the antenna is not vertical, a
trigonometric measurement error can occur with respect to the water. The
maximum range is reduced because of the off-axis return signal.
5. Orient the sensor such that one of its polarization markings is aligned
towards the wall or pier (see FIGURE 4-1 and TABLE 4-4).
3