Owner's manual
CS420-L and CS425-L Pressure Transducers 
    PortSet (1 ,1 )' Activate AM16/32 
    LoopCount = 1 'Reset loop counter 
 SubScan (0,Sec,4) 
 PulsePort (2 ,10000) 'Switch to next transducer, delay 10000 us before measurement 
 BrFull6W(WellDepth(LoopCount),1,mV2500,mV25,1, Vx1,1, 
 2500,True,True,0,250,Mult(LoopCount),Offset(LoopCount)) 
      LoopCount = LoopCount + 1 
 NextSubScan 
    PortSet (1 ,0)' Deactivate AM16/32 
 CallTable WellData 
 NextScan 
EndProg 
8.1.3.2 CR3000 Program 
'CR3000 
'For use with an AM16/32 in a 6 wire configuration, with four sensors. 
Public WellDepth(4) 
Dim LoopCount 
Dim Mult(4) 
Dim Offset(4) 
DataTable (WellData,1,-1) 
 DataInterval (0,30,Sec,10) 
 Sample (4,WellDepth(),FP2) 
 Maximum (4,WellDepth(),FP2,False,True) 
 Minimum (4,WellDepth(),FP2,False,True) 
EndTable 
BeginProg 
 Scan (5,Sec,0,0) 
    'Load multiplier and offset for each individual sensor 
 Mult(1) = 2.2975 
 Mult(2) = 1.0560 
 Mult(3) = 3.0001 
 Mult(4) = 1.9821 
 Offset(1) = .36781 
 Offset(2) = .4011 
 Offset(3) = .2103 
 Offset(4) = 1.0029 
    PortSet (1 ,1 )' Activate AM16/32 
    LoopCount = 1 'Reset loop counter 
 SubScan (0,Sec,4) 
 PulsePort (2 ,10000) 
17 