CS106 Barometric Pressure Sensor 
6.2  Accuracy 
Total Accuracy***:  ±0.3 mb @ +20 °C 
    ±0.6 mb @ 0 to +40 °C 
    ±1 mb @ –20 to +45 °C 
    ±1.5 mb @ –40 to +60 °C 
Linearity*:  ±0.25 mb @ 20 °C 
Hysteresis*:  ±0.03 mb @ 20 °C 
Repeatability*:  ±0.03 mb @ 20 °C 
Calibration Uncertainty**:  ±0.15 mb @ 20 °C 
Long-Term Stability:  ±0.1 mb per year 
*  Defined as ±2 standard deviation limits of end-point non-linearity, 
hysteresis error, or repeatability error 
**  Defined as ±2 standard deviation limits of inaccuracy of the working 
standard at 1000 mb in comparison to international standards (NIST) 
*** Defined as the root sum of the squares (RSS) of end-point non-linearity, 
hysteresis error, repeatability error and calibration uncertainty at room 
temperature 
6.3  General 
Dimensions:  9.7 x 6.8 x 2.8 cm (3.8 x 2.7 x 1.1 in) 
Weight:  90 g (3.2 oz) 
Housing Material:  ABS/PC blend 
Supply Voltage:  10 to 30 Vdc 
Supply Voltage Control:  When the internal jumper is closed, the CS106 
is on continually. When the jumper is open, 
the CS106 can be turned on/off with 5 Vdc/ 
0 Vdc. 
Supply Voltage Sensitivity:  negligible 
Current Consumption:  <4 mA (active); <1 µA (quiescent) 
Output Voltage:  0 to 2.5 Vdc 
Warm Up Time:  1 s 
Pressure Fitting:  barbed fitting for 1/8 in I.D. tubing 
Overpressure Limit:  2000 mb 