Owner's manual

Appendix A. Daily Average
08: Average (P71)
1: 1 Reps
2: 2* Loc [ P_const ]
09: If Flag/Port (P91)
1: 10 Do if Output Flag is High (Flag 0)
2: 30 Then Do
10: Z=X+F (P34)
1: 3* X Loc [ P_mb_avg ]
2: 800 F
3: 3* Z Loc [ P_mb_avg ]
11: End (P95)
12: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1440 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
13: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 10 Array ID
14: Real Time (P77)
1: 0110 Day,Hour/Minute
15: Sample (P70)
1: 1 Reps
2: 3* Loc [ P_mb_avg ]
16: Maximize (P73)
1: 1 Reps
2: 10 Value with Hr-Min
3: 1* Loc [ P_mb ]
17: Minimize (P74)
1: 1 Reps
2: 10 Value with Hr-Min
3: 1* Loc [ P_mb ]
-Input Locations-
1 P_mb
2 P_const
3 P_mb_avg
* Proper entries will vary with program and datalogger channel, and input location assignments.
** With a 21X, use the 5000 mV measurement range.
A-2