11: Z=X-Y (P35) ; Subtract zeroed value from measurement.
1: 1 -- X Loc [ mVPerVG01 ]
2: 17 -- Y Loc [ mVPerVZ01 ]
3: 33 Z Loc [ Vr_1 ]
12: Z=X*F (P37)
1: 33 X Loc [ Vr_1 ]
2: -2 F
3: 34 Z Loc [ One_2Vr ]
13: Z=X+F (P34)
1: 34 X Loc [ One_2Vr ]
2: 1000 F
3: 34 Z Loc [ One_2Vr ]
14: Z=X/Y (P38)
1: 33 X Loc [ Vr_1 ]
2: 34 Y Loc [ One_2Vr ]
3: 35 Z Loc [ Vr_1_2Vr ]
15: Z=X/Y (P38)
1: 35 X Loc [ Vr_1_2Vr ]
2: 36 -- Y Loc [ AdjGF01 ]
3: 52 -- Z Loc [ uStrain01 ]
16: Z=X*Y (P36)
1: 52 -- X Loc [ uStrain01 ]
2: 68 Y Loc [ Number4e3 ]
3: 52 -- Z Loc [ uStrain01 ]
17: End (P95)
18: If Flag/Port (P91)
1: 12 Do if Flag 2 is High
2: 30 Then Do
19: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1 Interval (same units as above)
3: 2 Call Subroutine 2 ; Outputs data to FinalStorage.
20: End (P95)
*Table 2 Program
02: 0
*Table 3 Subroutines
1: Beginning of Subroutine (P85) ; Measures ZeroOffset strain reading and sets GF
1: 1 Subroutine 1
2: Do (P86) ; Setup so Subroutine does not get called again.
1: 11 Set Flag 1 High
3: Z=F (P30) ; Lead Length Resistance per 100 feet.
1: 2.5 F ; 0.025 Ohms/Foot for 24 gauge copper stranded wire.
2: 0 Exponent of 10
3: 69 Z Loc [ LeadOhms ]
4: Bulk Load (P65) ; Load lead length of the gages in 100s of feet:
1: 3.0 F ; Gage01
2: 4.4 F ; Gage02
3: 8 F ; Gage03
4: 12 F ; Gage04
5: 14 F ; Gage05
6: 19 F ; Gage06
7: 15 F ; Gage07
8: 13 F ; Gage08
9: 70 Loc [ LeadFt01 ]