Specifications

If the meter does not display a message after
inserting a test strip.
PROBABLE CAUSE
WHAT TO DO
a. Battery exhausted Replace battery
b. Battery incorrectly
installed or absent.
Check that the battery is
correctly installed with the
positive “+” side up.
c. Test strip inserted
upside down or
incompletely.
Insert the test strip correctly
with the contact bars end first
and facing up.
d. Defective meter. Call Customer Service.
Problem Solving
If the test does not start after applying the sample:
PROBABLE CAUSE
WHAT TO DO
a. Insufficient blood
sample.
Repeat the test with a new
test strip and a larger sample.
b. Defective test strip.
Repeat the test with a new
test strip.
c. Sample applied after
automatic shutoff (2
minutes after last user
action).
Repeat the test with a new test
strip; apply sample only when
flashing drop of blood appears
on the display.
d. Defective meter.
Call Customer Service.