4.3.3 Change Output Values with the Knob in SET Mode
The SET mode of operation is a mode in which changes to output parameters made with the
Enter key is pressed. The AC source
he Set key twice. A blinking cursor indicates SET mode is
active.
knob or the entry keypad do not affect the output until the
is put in this SET mode by pressing t
To change the output voltage:
Counter
Clock
wise
Clock
wise
INCRDECR
1. Pre
e
OLT field to indicate a change in
FREQ entry
se
7. A blinking underline cursor will appear in the data for the FREQ field to indicate a change in
quency output values are now present at the output. The unit has
ode of operation until the Set key is pressed again.
ss the Set key twice
2. Place the cursor on the VOLT entry
3. Rotate the Voltage knob clockwise to increase the value, counterclockwise to decrease th
value
4. A blinking underline cursor will appear in the data for the V
settings but the output remains unchanged.
5. Place the cursor on the
6. Rotate the Frequency knob clockwise to increase the value, counterclockwise to decrea
the value
settings but the output remains unchanged.
8. Press the Enter key.
Both new voltage and fre
returned to immediate m
Note that output settings such as voltage and frequency can be changed from the measurement
screen as well. If all three phases are selected, slewing the Voltage knob will change the output
voltage on all three phases. If only one phase is selected, only the output of the selected phase
will be affected.