4.2.9.1 CALIBRATION Menus
routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.
This menu also contains the LCD viewing angle adjustment.
The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the
calibration screens, the calibration password must be entered first.
Note: Refer to chapter 6 for details on
ENTRY DESCRIPTION
VIEW ANGLE 0 -10 to +1 LCD viewing angle adjustment.
CAL PWORD V range Calibration password required to access all
is the high voltage range value. [270] The
password can be entered using the Voltage
calibration screens. The calibration password
shuttle followed by the ENTER key.
Measurement Calibration Screens
MVOLT F/S ± 9999 Calibration coefficient for full-scale voltage
measurement.
MCURR F/S ± 9999 Calibration coefficient for full-scale current
measurement.
Output Calibration Screens
VOLT F/S ± 9999 Calibration coefficient for full-scale voltage
output. (calibrated by measurement)
CURR F/S ± 9999 Calibration coefficient for full-scale current
ouptut. (calibrated by measurement)