4.2.8 REGISTERS Menu
non-voltage setup storage of the power source. A
in registers numbered from 0 through 15. Each
plete front panel setup, including the programmed
different setups and transient programs.
wer-on setting as assigned by the u
The registers menu provides access to the
total of 16 front panel setups can be stored
register except register 0 can hold the com
transient list. This allows for quick recall of
Register 0 is reserved to be used as the po ser. To have the
power source start in a specific setting, save the desired setting to Register 0 and assign
register zero as the power-on default in the C
can be set to power up w th the RST factory
ONFIGURATION menu. Alternatively, the FCS
default settings. See 4.9 for factory default settings. i
ENTRY DESCRIPTION
SAVE REG 0 – 15
-on state
Saves the selected setup and transient list from
memory. (Setup only for Reg 0) The left knob
(Voltage) may be used to scroll through the
available list of setup register numbers.
Use the ENTER key to perform the save
operation.
Register 0 can be assigned as the power
setup from the CONFIGURATION menu. A valid
setup must be saved in REG0 to do so.
Note that REG0 only saves the setup, not the
transient list. All other registers also save the
transient list.
RECALL REG 0 – 15
left knob
o.
Note that REG0 only saves the setup, not the
transient list. All other registers also save the
transient list.
Recalls the selected setup and transient list to
memory. (Setup only for Reg 0) The
(Voltage) may be used to scroll through the
available list of setup register numbers.
Use the ENTER key to perform the recall
operation.
Register 0 can be assigned as the power-on state
setup from the CONFIGURATION menu. A valid
setup must be saved in REG0 to do s
California Instruments 67