User manual

User Manual – Rev B FCS Series II
2.6.2 -EHV Option - Supplemental Specifications
Specifications for FCS units with -EHV voltage range option installed are same as standard FCS
unit except where noted below.
Frequency
Range: 45 Hz - 1000 Hz
Voltage:
Ranges (L-N):
Low range 0 - 200 Vrms
High range 0 - 400 Vrms
Current
Model FCS18-1-HV (single phase) FCS18-3-HV (three phase)
200 V Range. 90 Arms 30 Arms
400 V Range. 45 Arms 15 Arms
Note: Current derates linearly from 50% of voltage range to 20% of specified current at 10% of
voltage range.
Note: Current, maximum amps per phase available. For FCS36/2, currents are times two. For
FCS54/3, currents are times three. 35° C ambient max.
Repetitive Peak Current
200 V Range. 250 Apk 84 Apk
400 V Range. 125 Apk 42 Apk
California Instruments 23