User manual
User Manual – Rev B FCS Series II
SXF 106 Tests
Voltage distortion spectrum test require the use of additional external equipment. In particular, a
high bandwidth (100Khz) AC source, a high frequency coupling transformer and coup
networ as sho tandard. For this test, the AC s e
can be programmed to a steady state condition of 400 Hz and Vnominal of 115 Vrms or
Vrms.
ling
k wn in figure SXF106-1 of the Mil-Std 704 s o c
230
ur
d SXF-106 Test Setup.
SXF 107 Tests – R
Figure 11-12: Require
evision F
SXF107 Total Voltage Distortion Rev F
Condition Duration (s)
A 0.05 % VTHD 1800
F = 100%
H3 = 2.75%
H5 = 2.75%
H7 = 1.97%
H9 = 1.53%
H11 = 1.25%
H13 = 1.06%
H15 = 0.92%
SXF 108 Tests
SXF108 are DC offset tests and are not supported by the FCS Series AC Power sources.
Contact California Instruments for DC Supply information. (
sales@calinst.com )
California Instruments 187