User manual
User Manual – Rev B FCS Series II
TVF 103 Tests – Revision F
TVF103 Voltage Phase Difference - Rev F
Condition Duration (s)
A A = 0, B = 116, C = 240 1800
B A = 0, B = 124, C = 240 1800
TVF 104 Tests - Revision F
TVF104 Voltage Modulation - Rev B through F
Condition Duration (s)
Run at 400, Repeat at 360, 600 and 800 Hz
A Vmod F = 1.0 Hz, V = 0.375 Vrms 1800
B Vmod F = 1.7 Hz, V = 0.375 Vrms 1800
C Vmod F = 10 Hz, V = 2.5 Vrms 1800
D Vmod F = 25 Hz, V = 2.5 Vrms 1800
E Vmod F = 70 Hz, V = 0.375 Vrms 1800
F Vmod F = 100 Hz, V = 0.375 Vrms 1800
G Vmod F = 200 Hz, V = 0.375 Vrms 1800
TVF 10 R5 Tests – evision F
TVF105 Frequency Modulation Rev F
Condition Duration (s)
Run at 400, Repeat at 362, 600 and 798 Hz
A Hz 1800 Fmod F = 1 Hz/sec, Fmod Amp = 4 Hz± 2
B Hz 1800 Fmod F = 5 Hz/sec, Fmod Amp = 4 Hz± 2
C ± 2 Hz 1800 Fmod F = 10 Hz/sec, Fmod Amp = 4 Hz
D Fmod F = 25 Hz/sec, Fmod Amp = 4 Hz± 2 Hz 1800
E Fmod F = 100 Hz/sec, Fmod Amp = 4 Hz± 2 Hz 1800
California Instruments 179