User manual

User Manual – Rev B FCS Series II
2.1.3 AC Measurements
Measurement specifications apply to single chassis FCS Series II AC source in three-phase
mode. See notes for other models and configurations.
Parameter Range
Accuracy (±)
Resolution
1 Phase Mode 3 Phase Mode
Frequency
1
45.00-1200.0 Hz 0.1% ± 1 digit 0.01 Hz to 81.91 Hz
0.1 Hz to 819.1 Hz
1 Hz > 819.1 Hz
Phase
45.00 - 100.0 Hz
100.0 - 1200 Hz
0.5°
0. 1°
RMS Voltage 0 - 400 Volts 0.05% + 0.25V 0.01 Volt
RMS Current 0 - 150 Amps 0.1% + 0.45A 0.1% + 0.15A 0.001 Amp
Peak Current 0 - 250 Amps 0.2% + 1.5A 0.1% + 0.45A 0.001 Amp
Crest Factor 1.00 – 10.00 1.5 % 1.5 % 0.01
VA Power 0 - 6 KVA 0.3% + 15 VA 0.15% + 5 VA 1 VA
Real Power 0 - 6 KW 0.3% + 15 W 0.15% + 5 W 1 W
Power Factor 0.00 - 1.00 0.03 0.01 0.01
Note: Accuracy specifications are valid above 100 counts. For multi-chassis configurations, Current
and Power range and accuracy specifications are times the number of chassis.
Note: Frequency measurement specification valid for output > 20 Vrms.
Note: Crest Factor accuracy applies for Irms > 50% of max.
Note: Power Factor accuracy applies for PF > 0.5 and VA > 50% of max.
2.1.4 Harmonic Measurements
Harmonic measurement specifications apply to FCS18-3 with –ADV option in three-phase
mode. See notes for single-phase mode or FCS18–1 with –ADV option.
Parameter Range
Accuracy (±)
Resolution
Frequency fundamental 45.00 - 81.91 Hz
82.0 - 819.1 Hz
> 819.1 Hz
0.1% ± 1 digit 0.01 Hz
0.1 Hz
1 Hz
Frequency harmonics 45.00 Hz – 16 kHz 0.1% + 2 digits 0.1 Hz
Voltage
Fundamental 0 - 400 Volts 0.05% + 0.25V 0.01V
Harmonic 2 - 50 0.1% + 0.1%/kHz + 0.25 0.01V
Current
Fundamental 0 - 20 Amps 0.1% + 0.15A 0.01A
Harmonic 2 - 50 0.1% + 0.1%/kHz + 0.05A 0.01A
Note: Current range and accuracy specifications are times three in single-phase mode. For multi-
chassis configurations, current accuracy specifications are times the number of chassis. The harmonic
measurement bandwith increases to 48Khz in single-phase mode.
1
Frequency measurement specifications valid with output voltage of 30Vrms or higher. If output relay is open,
frequency measurement will return 0.0 Hz.
California Instruments 16