SAC 105 Tests – Revisions B, C, D
SAC105 Frequency Modulation Rev B, C, D
Condition Duration (s)
A Fmod F = 1 Hz/sec, Fmod Amp = 1800 ±5 Hz
B Fmod F = 5 Hz/sec, Fmod Amp = ±1.75 Hz 1800
C Fmod F = 10 Hz/sec, Fmod Amp = ±1.20 Hz 1800
D Fmod F = 25 Hz/sec, Fmod Amp = ±0.85 Hz 1800
E Fmod F = 100 Hz/sec, Fmod Amp = ±0.58 Hz 1800
SAC 105 Tests – Revisions A
SAC105 Frequency Modulation Rev A
Condition Duration (s)
A Fmod F = 1 Hz/sec, Fmod Amp = ±4 Hz 1800
B Fmod F = 5 Hz/sec, Fmod Amp = ±4 Hz 18 0 0
C Fmod F = 10 Hz/sec, Fmod Amp = ±4 Hz 1800
D 1800 Fmod F = 25 Hz/sec, Fmod Amp = ±4 Hz
SAC 106 Tests
of additional external equipment. In particular, a
C source, a high frequency coupling transformer and coupling
network as shown in figure SAC106-1 of the Mil-Std 704 standard. For this test, the AC source
can be programmed to a steady state condition of 400 Hz and Vnominal of 115 Vrms
Vrms.
Voltage distortion spectrum test require the use
high bandwidth (100Khz) A
or 230
Figure 11-8: Required SAC-106 Test Setup.
California Instruments 146