User manual

User Manual – Rev B FCS Series II
FREQUENCY TRANSIENTS (A(CF) only)
Test 1
Seq. No. Volt/Frequency Time
1 115/400 5 minutes
2 115/350 5 sec.
3 115/320 0.2 sec.
4 0/320 0.2 sec.
5 115V/400 10 sec.
Test 2
Seq. No. Volt/Frequency Time
1 115/400 5 minutes
2 115/480 0.2 sec.
3 115/440 5 sec.
4 0/440 0.2 sec.
5 115V/400 10 sec.
FREQUENCY TRANSIENTS (Group 1 only)
Seq. No. Frequency Time
1 400 5 minutes
2 480 5 sec.
3 400 10 sec.
4 320 5 sec.
5 400 10 sec.
Table 8-14: Abnormal Frequency Transient
e at 115V and will remain at this voltage through out the test except
he test will cycle the frequency three times as shown in Table 8-14.
1 and test2 for the A(CF) category are done in
This test will set the voltag
for the A(CF) category. T
Each repeat will start from sequence 2. Test
succession as a single test.
California Instruments 127