8.6 EMERGENCY TEST
From the DO160 MENU scroll to the EMERGENCY AC entry using the up and down cursor
keys. Press the ENTER key to select the EMERGENCY screens. The screen will appear as
shown in Figure 8-8.
F 8-8: mergency reens
The E SCREEN e following
1 VOLT FREQ MIN
igure E Sc
MERGENCY has th tests:
2 VOLT FREQ MAX
3 VOLT UNBALANCE
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.
VOLT FREQ MIN
Standard/Group RTCA A(CF) A(NF) A(WF)
1Ф 100 100 100 100 Voltage
3Ф 101.5 101.5 101.5 101.5
Frequency 360 360 360 360
Standard/Group Group1 Group2 Group3
1Ф 104 104 104 Voltage
3Ф 105.5 105.5 105.5
Frequency 360 360 360
Table 8-8: Emergency Voltage and Frequency Minimum
Standard/Group RTCA Group1 Group2 Group3
1Ф 122 122 122 122 Voltage
3Ф 120.5 120.5 120.5 120.5
Frequency 440 440 650 800
Standard/Group Group1 Group2 Group3
1Ф 122 122 122 Voltage
3Ф 120.5 120.5 120.5
Frequency 440 650 800
Table 8-9: Emergency Voltage and Frequency Maximum
This test will set the voltage and frequency to a level defined by Table 8-8. The test will last for
30 minutes. The test will be repeated using the voltage from Table 8-9 and frequency from
Table 8-8. The ← key (backspace) will terminate the test at any time.