User Manual – Rev B FCS Series II
T1
T2 T3
0 Volt
F1 F2
Test no.: 21 22 23 24 25 26
Standard: I II III IV V VI
T1 (ms) 50 50 100 100 200 200
F1 (Hz) 360 Fmax 360 Fmax 360 Fmax
F2 (Hz) Fmax 360 Fmax 360 Fmax 360
650 Hz for Group2/A(NF)
T3 = 5 msec
Figure 8-7: Power Interrupt for Group2/A(NF) and Group3/A(WF)
VOLTAGE SURGE
This test requires 160V output voltage. If the power source is set at the low voltage range, the
high voltage range will be selected before the test starts. At the end of the test, the power
source will be switched back to the low range automatically
Fmax =
Fmax = 800 Hz for Group3/A(WF)
T2 = 20 msec
Voltage Time
Seq. No. RTCA Group 1 Group 2 Group 3 ALL
1 115 115 115 115 5 Minute
2 160 160 160 170 30msec
3 115 115 115 115 5 Sec.
4 60 70 70 70 30msec
5 115 115 115 115 5 Sec.
Table 8-5: Normal VoltageSurge Sequence
The output voltage will follow the sequence in Table 8-5. The above sequence will repeat itself
three times. Each repeat will start from sequence two. RTCA and Group 1 will run at 400 Hz.
Group 2 and A(NF) will run at 360 Hz and 650 Hz. Group 3 and A(WF) will run at 360Hz and
800Hz. The frequency will return to the nominal setting when the test is completed. The key
(backspace) will terminate the test at any time.