User manual

User Manual – Rev B FCS Series II
2. Specifications
Specifications shown are valid over an ambient temperature range of 25 ± 5° C and apply after a
30 minute warm-up time. Unless otherwise noted, all specifications are per phase for sine wave
output into a resistive load. For three phase configurations or mode of operation, all
specifications are for Line to Neutral (L-N) and phase angle specifications are valid under
balanced load conditions only.
2.1 Electrical
2.1.1 Input
Parameter Specification
Line Voltage:
(3 phase, 3 wire
+ ground (PE))
Standard: 208 VAC ± 10 %
Optional: 240 VAC ± 10%
380 VAC ± 10%
415 VAC ± 10%
480 VAC ± 10%
Note: Each FCS chassis requires its
own AC service.
Line VA:
(total)
24 KVA. at nominal input voltage.
Line Current:
(per phase)
65 Arms @ 208 VAC
56 Arms @ 240 VAC
36 Arms @ 380 VAC
33 Arms @ 415 VAC
28 Arms @ 480 VAC
Currents shown are for single chassis models and per phase. For multi-
chassis configurations, currents are per chassis.
Line Frequency: 47- 63 Hz
Efficiency: 85 % (typical) depending on line and load
Power Factor: 0.85 (typical)
Inrush Current: 250 Apk max. Note: Each FCS chassis
requires its own AC service.
Hold-Up Time: > 10 ms
Isolation
Voltage:
400 VAC RMS input to output
1350 VAC input to chassis
California Instruments 12