Instruction manual
90
2. Depress the MNU key several times until the Menu screen is displayed as illustrated
below:
SNC = 01 RNG= 04
3. Enter the key sequence: 959 ENT.
4. Depress the MNU key several times until the configuration menu screen is
displayed.
CFG = 16 ALM = 17
FLM = 18 CLM = 19
5. Enter the key sequence 13 ENT to access the CAL VLT screen.
6. Example: If the maximum output voltage is 135 volts and if all phases of the
voltage measurement are to be calibrated to 135.0, enter the key sequence:
135 PRG ENT
After about 5 seconds, the volt measurement function will be calibrated for all three
phases.
If only one phase is to be calibrated to 135.0 volts (ex., phase A) enter the key
sequence:
135 A PRG ENT