4.3. PERIODIC CALIBRATION
The following periodic calibration adjustments should be performed on a 1 year interval.
4.3.1. VOLTAGE CALIBRATION
Apply power to the AC Power System and allow at least fifteen minutes for temperature
stabilization. Program the output to 60 Hz , 135 volts and perform the following steps:
1) Connect the AC voltmeter to the phase to be calibrated.
2) Depress the MNU key several times until the first menu screen is displayed as
illustrated below.
SNC = 01 RNG = 04
3) Enter the key sequence: 959 ENT
4) Depress the MNU key several times until the configuration menu screen is displayed
as illustrated below:
CFG = 16 ALM = 17
FLM = 18 CLM = 19
5) Select the CAL screen with the key sequence: 12 ENT
6) Depress the A, B or C key for the output phase to be calibrated. Verfiy that the
other phases are in the MON mode. The a, B, or C keys will switch the respective
phase between the MON and PRG mode. If the output voltage is greater than the
programmed output, depress and hold the key until the correct output volrage has
been reached. The key is used if the output is less than the programmed value.
7) Repeat step 6 for other outputs to be calibrated..
4.3.2. VOLTAGE MEASUREMENT CALIBRATION
For calibration of voltage measurement first perform the output voltage calibration and then
perform the followings steps:
1. Remove the load from the AC Power System and program 60 Hz and the 135.0
volts.