Instruction manual
83
4. CALIBRATION PROCEDURE
4.1. GENERAL
The calibration is divided into two categories; a periodic and a nonperiodic calibration. The
periodic calibration should be performed at a 1 year interval. The nonperiodic calibration should
only be performed if the periodic calibration cannot be performed or if an adjustable subassembly
is replaced.
The following is a listing of paragraphs that may be performed to fix an indicated problem. Any
AC Power System with a 1-phase output or that has more than one chassis will have paralleled
amplifiers.
PARAGRAPH TITLE
4.3.1 OUTPUT VOLTAGE CALIBRATION
This is a periodic calibration of the output voltage.
4.3.2 VOLTAGE MEASUREMENT CALIBRATION
This is a periodic calibration of the voltage measurement
function.
4.3.3 CURRENT MEASUREMENT CALIBRATION
This is a periodic calibration of the current measurement
function.
4.3.4 POWER MEASUREMENT CALIBRATION
This is a periodic calibration of the power measurement
function.
4.3.5 REMOTE MEASUREMENT CALIBRATION
4.4.1 OUTPUT FREQUENCY CALIBRATION
This is a nonperiodic calibration of the output frequency.
4.4.2 AMPLIFIER GAIN ADJUSTMENT
These are nonperiodic adjustments of the output amplifier
gain. The adjustments may have to be performed if there is an
OUTPUT FAULT error message.
4.4.3 OUTPHASE ANGLE CALIBRATION
This is a nonperiodic calibration. The calibration is required if
there is an error in the phase B or C output phase angle
relative to phase A.