Table 3-15: Status Byte Values
SRQ
1 0
REPORTED MESSAGE CAUSE
STATUS BYTE
64 0
65 1
66 2
67 3
68 4
69 5
70 6
72 8
90 26
91 27
92 28
93 29
94 30
95 31
96 32
97 33
98 34
99 35
100 36
63
40
OUTPUT A FAULT
OUTPUT B FAULT
OUTPUT AB FAULT
OUTPUT C FAULT
OUTPUT AC FAULT
OUTPUT BC FAULT
OUTPUT ABC FAULT
MODULE FAILURE
RNG RANGE ERROR
AMP RANGE ERROR
FRQ RANGE ERROR
PHZ RANGE ERROR
CRL RANGE ERROR
RMPA RANGE ERROR
SYNTAX ERROR
BUS LOCAL ERROR
EXT SYNC ERROR
CPU MEMORY FAULT
DMA OVERFLOW ERROR
STA OK
Overload or sense line fault
Overload or sense line fault
Overload or sense line fault
Overload or sense line fault
Overload or sense line fault
Overload or sense line fault
Overload or sense line fault
Amplifier failure
RNG value greater than highest range
AMP value greater than RNG value
FRQ value is less than 45 or greater than 550.0
PHZ value greater than ±999.0
CRL value greater than maximum value
DLY, STP or VAL values wrong
Wrong string SYNTAX
Remote message sent while in local mode
No external sync input or signal not between 45
and 550 Hz
CPU failed self-test
Remote message greater than 256 bytes
The response after SRQ2 is included in a setup
string and the execution of the string or
measurement is completed.
No problems