Instruction manual

71
The following example will program the default frequency to 400 Hz.
FLM A 400
3.7.5.14.TO PROGRAM THE DEFAULT OUTPUT VOLTAGE (INI A)
The default voltage is the output voltage after power-up, IEEE-488 Device Clear or an
Amplitude fault.
The following example will program the default voltage to 5 volts.
INI A 5
NOTE: If the default voltage is programmed to a value less than 5 volts, the
settling time will increase. In addition, there may be an amplitude fault
when the voltage is programmed from the default voltage to a value less
than 50% of full scale.
3.7.5.15.DEFAULT CURRENT LIMIT (INI C)
The default current limit is the value after power-up or IEEE-488 Device Clear.
The following example will program the default current limit to 10 amps.
INI C 10
3.7.5.16. 1 OR 3 PHASE MODE (MOD)
The MOD header with the required PHS extension is used to identify the mode command. The
argument is either a 1 or a 3 to specify the 1-phase or 3-phase mode respectively. A mode
change by the AC Power System will cause the output voltage to go to the default value.
CAUTION: Do not program the Phase Mode without first configuring the output
wiring. Refer to Section 2. If the MODE Option is installed, relays will
automatically change the output wiring between 1-phase and 3-phase
configurations.
The following string will set the AC Power System to the 3-phase mode:
MOD PHS 3