To set any of the default values, perform the following steps:
1. Depress the MNU key several times until the first menu screen is displayed as
illustrated below:
SNC = 01 RNG = 04
2. Enter the key sequence: 9 5 9 ENT
3. Depress the MNU key several times until the configuration menu screen is
displayed as illustrated below:
CFG = 16 ALM = 17
FLM = 18 CLM = 19
To program the default frequency enter the key sequence:
1 8 ENT A
Next enter the default frequency followed by depressing the PRG and ENT
key.
To program the default voltage range, perform step 1 through 3. Next enter the key
sequence:
17 ENT A
At this point if 0 is entered the default voltage range is the Low Range.
If an 8 is entered the default voltage range will be the High Range.
To make High Range the default, continue the key sequence with:
8 PRG ENT
To program the default voltage, perform steps 1 through 3. Next enter the key sequence:
2 9 ENT A
At this point the default voltage from 0 to 5 may be entered. If a value of less
than 5 volts is entered, the output may fault when a voltage is programmed to
a value that is less than 50% of full scale.
To make 5 volts the default voltage, continue the key sequence with:
5 PRG ENT