47
5. Store this program in Register 1:
1 REG ENT
The second portion of the program will be stored in Register 2.
6. Select the FRQ screen and program 62 Hz:
6 ENT 62 PRG
7. Select the AMP screen and program 135 volts:
5 ENT 135 PRG
8. Select the RMP screen and program DLY = 0.01, STP = 0.1 and VAL = 115:
9. Store this program in Register 2:
2 REG ENT
To initiate the program:
1 REC ENT
3.5.3.9.SIMULTANEOUS RAMPS
Two outputs may be simultaneously ramped or stepped by enabling two parameter screens.
The screen first selected will be the dependent parameter. The last parameter screen selected
before entering the ramp (RMP) screen is the independent parameter. The independent
parameter is used to specify the number of steps in a ramp. Since the dependent parameter
has as many steps as the independent parameter, the step (STP) size must be calculated so
the dependent parameter will not exceed its maximum value. The following example will
specify frequency as the independent parameter and voltage amplitude as the dependent
parameter. Refer to paragraph 3.5.8 for more information.
The following example will ramp frequency from 360 to 440 Hz at a rate of .2 Hz per .2
second, while each .2 Hz step causes the amplitude to go from 10 volts to 210 volts in .5 volt
steps.
1. Select the AMP screen and specify the starting voltage of 10 volts:
5 ENT 10 PRG
2. Select the FRQ screen and specify the starting frequency of 360 Hz.
6 ENT 360 PRG