Instruction manual
41
Program the output to 60.23 hertz:
60.23 PRG ENT
To incrementally increase the output frequency to a desired value:
↑ (Hold until desired frequency is reached.)
The output frequency may be programmed down to 17 Hz if the programmed voltage
amplitude (AMP)is less than the full scale voltage range. The low frequency limit can be
determined from the following formula:
FRQ = 45 * (AMP)/Voltage Range
3.5.3.4.PHASE ANGLE (PHZ=7)
Select the Phase (PHZ) screen by entering:
7 ENT
Program phase C to .5 degree relative the phase A:
0.5 C PRG ENT
To enable the External Sync input and to program phase A to 90.0 degrees relative to the
External Sync input, perform the following sequence:
1. Select the SNC screen:
1 ENT
2. Program EXT:
EXT PRG
At this point the phase screen must be accessed by repeatedly depressing either the MON or
PRG keys. After the Phase (PHZ) screen is displayed, enter the following key sequence to
simultaneously enable the External Sync and program phase A to 90 degrees:
9 0 A PRG ENT
The up (↑ and down (↓) keys may be used to increment or decrement any or all output
phases.