Instruction manual
17
6) Apply full loads to each phase. Verify that the voltage of each phase remains
within 0.135 volts of the no-load voltage. The waveforms shall still appear
clean on the oscilloscope and have less than 2% distortion. (Full load is 3.0A
per 6KVA phase at 135 volts output.)
7) Program the Current Limit to 25.0 amps:
8 ENT
25 PRG ENT
8) The display should show the error message 'CRL FAULT'. The output will
default to a programmed value of 0.0 volts.
If the FCS 18 has the PRC Option installed, perform the following steps:
9) With the front panel keypad, program the HI output voltage range with the
following key sequences:
4 ENT To select the Range screen (RNG)
To program the HI range (ex., 270 volt range):
270 PRG ENT
10) Program the output to maximum voltage with the following key sequences:
Depress the MON key 1 time to select the Amplitude screen (AMP).
To program the voltage to maximum.
(Maximum Range value) PRG ENT
Verify that the front panel voltmeter indicates approximately the programmed
value for all three phases or for phase A for a 1-phase system.