Instruction manual
6
* Denotes guaranteed specifications. All other specifications are
supplementary.
Protection
Input Under Voltage/Phase Loss Yes
Sense Line Fault Open, short, reversal
Temperature Shutdown, no automatic reset
Incorrect Signal Frequency Shutdown, no automatic reset
Measurements (23°
°°
°C ±5°
°°
°C)
Range Resolution Accuracy
* Voltage 0 to 312V 0.1 volt ±10 digits of resolution
0 to 200V 0.1 volt ±20 digit
(HV4 Option 200 to 400V 0.1 volt ±35 digits
* Current 133 0.1 Amp ±10 digits of resolution
1333 1 Amp ±10 digits of resolution
* Power 18kW 0.01KW ±10 digits of resolution
180kW 0.1 KW ±10 digits of resolution
* Frequency 45 to 99.99 0.01 ±0.02 Hz
100 to 499.9 0.1 ±0.2 Hz
500 to 999.9 0.1 ±0.5 Hz
1000 to 1200 1 ±10 Hz
* Phase Angle 0 to 360 0.1 ±2°
NOTE: For frequencies below 45 Hz, add 2% of measured value to measurement.
Control
Front Panel Controls Circuit Breaker
Keypad 20 button (2 row by 10)
Bus IEEE-488.1
Subsets SH1, AH1, T6, L3, SR1, RL2, DC1, DT1,
PPO, CO
Data Transfer Rate 200K Bytes/Sec.
Language APE
Function:
Range
Resolution Accuracy Initialization
* Voltage 0-135V 0.1 volts 0.2% of range 5
-156V 0.1 volts 0.2% of range 5
5-200 0.1 volts ±1.5% of range 5
HV4 Option 200-400 0.1 volts ±3.0% volts 5
NOTE: Add 2% of programmed value below 45 Hz.
* Frequency 17-99.99 0.1% ±0.005% 60 Hz
100-999.9 0.1. ±0.005% 60 Hz