Instruction manual

93
(Measured phase C current) C PRG ENT
4.3.4. POWER MEASUREMENT CALIBRATION
For calibration of power measurement perform the following steps:
1. Program 60 Hz and 135.0 volts and the maximum Current Limit value.
2. If any calibration screen is already displayed, the Power Calibration screen (CAL
PWR) may be displayed by depressing either the MON or PRG keys and then skip
to step 6. If a calibration screen is not displayed, depress the MNU key several
times until the screen shown below is displayed:
SNC = 01
3. Enter the key sequence: 959 ENT
4. Depress the MNU key several times until the configuration menu screen is displayed
as shown below:
CFG = 16 ALM = 17
FLM = 18 CLM = 19
5. Enter the key sequence 15 ENT to access the CAL PWR screen.
6. Apply the load resistance value from Table 4-1 to the phase A output only. Measure
the phase A output current with the AC Digital Voltmeter and Current Transformer.
Measure the phase A voltage from the phase A sense to neutral sense. Multiply the
voltage and current values to determine the power value in kilowatts. (Note: The
load must be resistive for the correct power value.)
7. On the keypad enter the key sequence:
(Measured phase A power) A PRG ENT
8. Repeat step 6 for phase B.
9. Repeat step 7 except enter the key sequence:
(Measured phase B power) B PRG ENT
10. Repeat step 6 for phase C.
11. Repeat step 7 except enter the key sequence: