5.6 Additional Setup Parameters for Harmonics
If this is first time using the CTS 3.0 system, it may be necessary to check the Additional Setup
settings. These settings can be displayed by clicking on the "Additional Setup" button on the right
hand side of the main setup window.
Figure 5-4: Additonal Settings for Harmonics Test
The following settings are available from this dialog window:
Setting Description
Pre test time This is the time during which to apply power to the EUT before staring the
harmonics measurements. This capability may be used to skip any startup
behavior of the EUT that may affect the outcome of the test. High EUT inrush
currents that exceed the AC Source's maximum peak current capability can result
in high voltage distortion. This would invalidate the test. The Pre-test time setting
can be used to hold off measurements until after the EUT has reached a steady
state condition.
Source Control The CTS system is normally supplied with a suitable AC power source to provide
stable and low distortion AC power to the EUT. The Source Control section of the
additional setup dialog is used to configure the CTS system for the AC Source
used. Available options are:
Use the AC source
from its front panel
only
This option may be needed if the AC Line is used or the AC
source used does not offer remote control capabilities. This
way of operating puts more responsibility on the operator to
ensure the correct power is applied during the test.
RS232C control Used with AC Sources that have RS232C control
capability. The Verify Source must be used to make sure
the AC Source can be found. The AC source must be
connected using the serial cable provided and must be
powered up. The View Bus button may be used to bring up
the command trace for trouble shooting purposes if the
presence of source cannot be verified. When using remote
control, the user can select to have the front panel locked
or not by setting the Disable checkbox.
Note that su
pp
ort for the Pacific Power Sou
r
ce AM
X
with -