User manual

Compliance Test System 3.0 User Manual
Measurement Specification Unit
Power
Range PACS-1 / PACS-3 0.1 - 12,500 W/ph
Range PACS-3-75 0.1 - 22,500 W/ph
Accuracy ± 0.25 % ± 0.25 % FS ± 20mW mW
Resolution 0.1 W
Apparent Power
Range PACS-1 / PACS-3 0.01 - 12,500 VA/ph
Range PACS-3-75 0.01 - 22,500 VA/ph
Accuracy ± 0.15% ± 0.15% FS ± 20 mVA
Resolution 0.01 VA
Power Factor
Range 0.00 - ± 1.00
Accuracy ± 0.05
Resolution 0.01
Crest Factor
Range 20:1
Accuracy ± 0.005
Resolution 0.001
Frequency
Range 45 - 65 Hz
Accuracy 0.1 % of reading Hz
Resolution 0.1 Hz
Harmonic Analysis
Range Fundamental to 40
th
Accuracy Fundamental ±0.05% FS±0.05%/kHz
Accuracy Harmonics ±0.1 %±0.1%/kHz
Measurement window 10, 12 or 16 periods
Smoothing filter 1.5 sec
Flicker
Pst Range 0.1 - 10 Pst
Accuracy 3 %
Resolution 0.01
Integration time 10 min
Plt Range 0.1 - 10
d
max
Range 0 - 100 %
d
c
Range 0.1 - 100 %
d
t
Range 0.1 - 100 %
d
t
over3%(4%) Range 0 - 1000 ms
144 Revision N California Instruments