Compliance Test System 3.0 User Manual
User Prompts
Some EUT's will require resetting and readjustment when the power dips or is interrupted for a
long period. If this is the case, then set GapDelay0 = -1. This will cause the test sequence to halt
and wait for user intervention. When ready to resume test press “OK”.
14.2.2 IEC413.413 File
[IEC-1000-4-13] IEC 61000-4-13 section header
Parameters that apply only to Flat Top Curve
Fcurv_lev = 90 Flat Curve Level at 90%
Fcurv_dwe = 120 Flat Curve Dwell time in seconds
Fcurv_pau = 5 Flat Curve Pause in seconds
Parameters that apply only to Over Swing Curve
Oswcurv_dwe = 120 Over Swing Curve Dwell time in seconds
Oswcurv_pau = 5 Over Swing Curve Pause in seconds
Oswcurv3_lev = 6 Over Swing Curve of third harmonic level at 6%
Oswcurv3_phs = 180 Over Swing Curve of third harmonic phase test of 180 degrees
Oswcurv5_lev = 4 Over Swing Curve of fifth harmonic level at 4%
Oswcurv5_phs = 0 Over Swing Curve of fifth harmonic phase test of 0 degree
Parameters that apply to Frequency Sweep
Swefreq1_lev = 3 Sweep Frequency Level of first range at 3%
Swefreq2_lev = 10 Sweep Frequency Level of second range at 10%
Swefreq3_lev = 5 Sweep Frequency Level of third range at 5%
Swefreq4_lev = 5 Sweep Frequency Level of fourth range at 5%
Swefreq5_lev = 2 Sweep Frequency Level of fifth range at 2%
Swefreq_dwe = 300 Sweep Frequency Dwell time in seconds
Swefreq1_stp = 5 Sweep Frequency Step size of first range in Hz
Swefreq2_stp = 10 Sweep Frequency Step size of second range in Hz
Swefreq3_stp = 10 Sweep Frequency Step size of third range in Hz
Swefreq4_stp = 10 Sweep Frequency Step size of fourth range in Hz
Swefreq5_stp = 25 Sweep Frequency Step size of fifth range in Hz
Parameters that apply only to Individual Harmonics
Indharm2_lev = 3 Individual harmonics Level for nth (second) harmonic in %
Indharm2_phs = 0 Individual harmonics Phase for nth (second) harmonic in degree
Indharm_dwe = 5 Individual harmonics Dwell time in second
Indharm_pau = 5 Individual harmonics Pause in second
Parameters that apply only to Interharmonics
Interharm1_lev = 2.5 Interharmonics Level for first range at 2.5%
Interharm2_lev = 5 Interharmonics Level for second range at 5%
Interharm3_lev = 3.5 Interharmonics Level for third range at 3.5%
Interharm4_lev = 2 Interharmonics Level for fourth range at 2%
Interharm5_lev = 1.5 Interharmonics Level for fifth range at 1.5%
Interharm_dwe = 5 Interharmonics Dwell time in second
Interharm_pau = 1 Interharmonics Pause in second