1.2.1 AC Source Models
The CTS system is supplied with different programmable AC power sources depending on the
configuration. The following AC sources are supplied with each CTS system:
Table 1-1: CTS AC Source Models
Model VA Power AC Source IEC
61000-4
PACS
model
Single Phase Systems
100-CTS AC Line none - PACS-1
1251RP-CTS 1250 VA 1251RP - PACS-1RP
3001iX-CTS 3000 VA 3001iX
PACS-1
5001iX-CTS (-400) 5000 VA 5001iX (-400)
PACS-1
10001iX-CTS (-400) 10000 VA 10001iX (-400)
(See note below)
PACS-1-75
Three Phase Systems
300-CTS-75 AC Line none - PACS-3-75
15003iX-CTS (-400) 15000 VA 15003iX
PACS-3
30003iX-CTS (-400) 15000 VA 30003iX-400
(See note below)
PACS-3-75
If you plan to use the AC line and are using a PACS-1 or PACS-3 system, you can skip this
paragraph as there is no AC source included with the system. If you are using any of the other
CTS configurations, the AC source includes its own user manual. We recommend you familiarize
yourself with the operation of the AC source and the included Windows Graphical User Interface
(CIGUI32 or CIGuiSII) for it before you start using the CTS system.
Note: PACS-1 and PACS-3 based CTS systems are rated for a maximum current of 40 amps
rms per phase. Higher current versions, PACS-1-75 and PACS-3-75, are available.
Existing 10001iX-CTS and 30003iX-CTS systems may not have been shipped with these
-75 model versions however. Damage to the standard PACS units can occur if higher
currents are present. Care must be taken to not exceed this rating when using the
system directly with the utility line or with the higher power sources. When using a
10001iX-CTS or 30003iX-CTS system, the lower voltage/higher current output range
should not be used unless the AC source’s current limit has been set to 40 amps or less
or the system was supplied with the -75 version of the PACS.