The above listed models are intended for use in the electromagnetic environment specified below. The customer or the
user of the above listed models should assure that they are used in such an environment.
Input AC power PORT
Immunity test Test Condition IEC 60601 Compliance
level
Electromagnetic
environment - guidance
Electrical fast
transient/bursts
IEC 610004-4
± 2 kV
100kHz Repetition frequency
± 2 kV Mains power quality should
be that of a professional
healthcare facility
environment and Home
healthcare environment.
Criteria B
Surges
IEC 61000-4-5
± 0,5kV, ±1 kV line(s) to line(s)
± 0,5kV, ±1 kV , ± 2 kV line(s) to
ground (a)
± 1 kV , Differential mode
± 2 kV Common mode
Mains power quality should
be that of a professional
healthcare facility
environment and Home
healthcare environment.
Criteria B
Conducted RF induced by
RF fields
IEC 61000-4-6
10 Vrms 150 kHz - 80 MHz
also ISM and Amateur Radio Bands
c)
80% AM 1kHz
10 Vrms Mains power quality should
be that of a professional
healthcare facility
environment and Home
healthcare environment.
Criteria A
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% UT;
0°,45°,90°,135°,180°,225°,270°,315°
0% UT;
0% UT;
70%
0% UT;
0%
0,5 Cycles
1Cycle
25/30 Cycles (50/60Hz)
250/300 Cycles (50/60Hz)
(5s)
Mains power quality should
be that of a professional
healthcare facility
environment and Home
healthcare environment.
.
If the user of the above
listed models requires
continued operation during
power mains interruptions,
it is recommended that the
above listed models are
powered from an
uninterruptible power
supply or battery.
Criteria A (max. Mains
voltage)
Criteria B (min. Mains
voltage)
Criteria A (max. Mains
voltage)
Criteria B (min. Mains
voltage)
Criteria A (max. Mains
voltage)
Criteria B (min. Mains
voltage)
Criteria B (max. Mains
voltage)
Criteria B (min. Mains
voltage)
Comment:
a) Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS. b) The ISM (industrial, scientific and medical) bands between 0,15 MHz
and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur
radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz
to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and
50,0 MHz to 54,0 MHz.