User's Manual

CONFIDENTIAL
CADI SmartSense – SmartNode SMN-822W User Manual
T-SMN-822-I02-A01
14
The above listed models are intended for use in the electromagnetic environment specified below. The customer or the user of the
above listed models should assure that they are used in such an environment.
Enclosure Port
Immunity test
Test Condition
IEC 60601
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV Contact ± 2,4,8,15 kV
Air
±8 kV Contact ± 15
kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Criteria B
Radiated RF EM fields
and Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80% AM 1kHz
10 V/m 80 MHz – 2,7
GHz
Mains power quality should be that of a
professional healthcare facility
environment and Home healthcare
environment.
Criteria A
385MHz (18Hz Pulse
Modulation)
27 V/m
450MHz (FM+/-5KHz deviation
1kHz sine or 18Hz Pulse
Modulation)
28 V/m
710MHz (217Hz PM) 9 V/m
745MHz (217Hz PM) 9 V/m
780MHz (217Hz PM) 9 V/m
810MHz (18Hz PM) 28 V/m
870MHz (18Hz PM) 28 V/m
930MHz (18Hz PM) 28 V/m
1720MHz (217Hz PM) 28 V/m
1845MHz (217Hz PM) 28 V/m
1970MHz (217Hz PM) 28 V/m
2450MHz (217Hz PM) 28 V/m
5240MHz (217Hz PM) 9 V/m
5500MHz (217Hz PM) 9 V/m
5785MHz (217Hz PM) 9 V/m
RATED power
frequency
magnetic fields
IEC 61000-4-8
50Hz or 60Hz 30 A/m Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Criteria A
Comment