User's Manual

EID-2 Optical Impression Device Operation Manual
4 Publication SD08300-A (Rev. 1.10)
WARNINGS
POWER
An internal battery backup unit will supply AC power to the system in the event of a power shortage.
Therefore, power is present in the internal hardware components even when the system is unplugged
from the wall. To ensure that power is removed from all components, follow these instructions:
Close all files and applications.
Press the front POWER switch to shutdown the system.
Press the MAIN POWER SHUT OFF switch on the back of the cart.
Unplug power cable from the wall socket.
To restore power, plug in the power cable and press the MAIN POWER SHUT OFF switch. The
system will beep to indicate return of power.
According to Department of Transportation (DOT) regulations, the system must be shipped with
internal battery leads disconnected. Please note that battery leads should be disconnected and
reconnected by authorized Cadent technicians only.
Power is supplied to the system via an isolation transformer.
WIRELESS LAN
The system comes equipped with a Wireless LAN unit. Using standard wired communications will
violate the AC mains power isolation.
SCANNER
Scanner emits red laser light (660 nm Class 1) as well as a multi-colored LED light. Normal usage of
the scanner does not present any danger to the eye. However, the doctor should refrain from shining
the light directly into the patients eyes.
The scanner cable should be handled with care. Avoid stepping on the scanner cable, rolling over the
cable with the cart wheels, knotting, or over-twisting the scanner cable.
The scanner head contains fragile parts and should be handled with care. Before starting to work, be
sure to read the instructions in Chapter 3 for disinfecting the scanner and replacing the disposable
sleeve.
GENERAL
Be careful not to block any of the air vents on the cart.
Refrain from opening any closed components. Except for the replacement of batteries in the mouse
and foot pedal, all closed components should be serviced by authorized Cadent technicians only.
The system is intended for indoor usage only and should not be exposed to direct sunlight.
Due to its weight, the system should be lifted and loaded by two or more persons.